Full scanners comparison |
|
Technology |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
System architecture |
analog |
analog |
hybrid |
hybrid |
hybrid |
digital |
|
Imaging modes |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
B, B+B, 4B, B+M, M |
yes |
yes |
yes |
yes |
yes |
yes |
|
Color Doppler (CFM) |
no |
no |
yes |
yes |
yes |
yes |
|
Power Doppler (PDI) |
no |
no |
yes |
yes |
yes |
yes |
|
Directional Power Doppler (DPDI) |
no |
no |
yes |
yes |
yes |
yes |
|
Pulsed Wave Doppler (PWD) |
no |
no |
yes |
yes |
yes |
yes |
|
Continuous Wave (CW) |
no |
no |
no |
no |
no |
yes |
|
B+PWD (Duplex) |
no |
no |
yes |
yes |
yes |
yes |
|
B+CFM/PDI/DPDI+PWD (Triplex) |
no |
no |
yes |
yes |
yes |
yes |
|
High Pulse Repetition Frequency (HPRF) |
no |
no |
no |
yes |
yes |
yes |
|
Ultrasound Technologies |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
Spatial Compound |
no |
no |
no |
2 frames, 10° |
2 frames, 10° |
2/3 frames, up to 25° |
|
B-steer for linear probes |
no |
no |
no |
yes, -10°/+10° |
yes, -10/+10 |
yes, up to -25°/+25° |
|
Trapezoid Imaging for linear probes |
no |
no |
no |
no |
no |
yes, up to 25° |
|
Tissue Harmonic Imaging (THI) |
no |
no |
no |
yes |
yes |
yes |
|
Advanced Speckle Reduction Imaging NeatView |
yes |
yes |
yes |
yes |
yes |
yes |
|
Scanning method |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
electronic linear, convex, microconvex |
yes |
yes |
yes |
yes |
yes |
yes |
|
phased array |
no |
no |
no |
no |
no |
yes |
|
scanning depth, cm |
3-24 |
3-24 |
2-30 |
2-30 |
2-30 |
2-31 |
|
Transducers |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
automatic transducer recognition |
yes |
yes |
yes |
yes |
yes |
yes |
|
ultra wide band multifrequency support |
yes |
yes |
yes |
yes |
yes |
yes |
|
supported frequency range, MHz |
2.0-8.0 |
2.0-10.0 |
2.0-10.0 |
2.0-12.0 |
2.0-14.0 |
2.0-16.0 |
|
supported transducers |
see here |
|||||
|
Focusing |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
transmit focusing |
variable |
variable, 8 zones |
variable, 8 zones |
variable, 8 zones |
variable, 8 zones |
variable, 8 zones |
|
receive focusing |
variable |
dynamic, 8 zones |
dynamic, 8 zones |
dynamic, 8 zones |
dynamic, 8 zones |
point to point, dynamic |
|
programmable focus area presets |
yes |
yes |
yes |
yes |
yes |
yes |
|
Automatic Image Optimization (single click) |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
B-image: gain, dynamic range, TGC sliders |
no |
no |
yes |
yes |
yes |
yes |
|
Color Doppler: CFM/PDI/DPDI gain |
no |
no |
yes |
yes |
yes |
yes |
|
Pulsed Waved Doppler: baseline, invert, PRF |
no |
no |
yes |
yes |
yes |
yes |
|
Ultrasound imaging |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
ultrasound image size: automatically adjustable to screen resolution |
yes |
yes |
yes |
yes |
yes |
yes |
|
gray scale: 256 |
yes |
yes |
yes |
yes |
yes |
yes |
|
color scale: 256 |
no |
no |
yes |
yes |
yes |
yes |
|
full motion and full size real-time ultrasound imaging, up to 120 fps* |
yes |
yes |
yes |
yes |
yes |
yes |
|
zoom mode: from 60% to 600% in scanning and freeze modes |
yes |
yes |
yes |
yes |
yes |
yes |
|
viewing area variable for frame rate maximizing: 6 steps |
yes |
yes |
yes |
yes |
yes |
yes |
|
thumbnail mode: up to 32 images |
yes |
yes |
yes |
yes |
yes |
yes |
|
unlimited programmable presets for clinically specific imaging |
yes |
yes |
yes |
yes |
yes |
yes |
|
cineloop size: thousands of frames (depends on installed memory size) |
yes |
yes |
yes |
yes |
yes |
yes |
|
Processing |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
Line Density selection for better resolution |
yes |
yes |
yes |
yes |
yes |
yes |
|
TGC Control, 5-10 sliders (customizable) 40 dB |
yes |
yes |
yes |
yes |
yes |
yes |
|
dynamic range |
yes |
yes |
yes |
yes |
yes |
yes |
|
overall gain control |
yes |
yes |
yes |
yes |
yes |
yes |
|
M - mode sweep speed control |
yes |
yes |
yes |
yes |
yes |
yes |
|
acoustic power control |
yes |
yes |
yes |
yes |
yes |
yes |
|
variable frame averaging |
yes |
yes |
yes |
yes |
yes |
yes |
|
brightness, contrast, gamma |
yes |
yes |
yes |
yes |
yes |
yes |
|
advanced gamma control: 8 fixed curves, 8 user defined (customizable) |
yes |
yes |
yes |
yes |
yes |
yes |
|
scan direction, rotation, up-down controls |
yes |
yes |
yes |
yes |
yes |
yes |
|
negative / positive control |
yes |
yes |
yes |
yes |
yes |
yes |
|
bi-linear interpolation |
yes |
yes |
yes |
yes |
yes |
yes |
|
echo enhancement control |
yes |
yes |
yes |
yes |
yes |
yes |
|
noise rejection function |
yes |
yes |
yes |
yes |
yes |
yes |
|
Color Doppler (CFM, PDI, DPDI) |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
PRF variable, kHz |
no |
no |
0.5-10 |
0.5-10 |
0.5-10 |
0.5-15 |
|
wall filter settings: 3 steps (5%, %10%, 15% PRF) |
no |
no |
yes |
yes |
yes |
yes |
|
gain control: 50 dB |
no |
no |
yes |
yes |
yes |
yes |
|
angle steering for linear transducers |
no |
no |
±10° |
±10° |
±10° |
±30° |
|
real-time spatial filter: 4 values |
no |
no |
yes |
yes |
yes |
yes |
|
CFM palette: 10 maps |
no |
no |
yes |
yes |
yes |
yes |
|
PDI palette: 11 maps |
no |
no |
yes |
yes |
yes |
yes |
|
B/Color priority control |
no |
no |
yes |
yes |
yes |
yes |
|
color threshold control |
no |
no |
yes |
yes |
yes |
yes |
|
CFM baseline control |
no |
no |
yes |
yes |
yes |
yes |
|
Doppler frequency selection: 2 frequencies / each transducer |
no |
no |
yes |
yes |
yes |
yes |
|
color frame averaging: 8 values |
no |
no |
yes |
yes |
yes |
yes |
|
Transparent Color Mapping (TCM): 10 values |
no |
no |
yes |
yes |
yes |
yes |
|
Pulsed Wave Doppler |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
PRF variable, kHz |
no |
no |
1-10 |
1-15 |
1-15 |
1-20 |
|
wall filter settings: 16 steps (2.5%-20% PRF) |
no |
no |
yes |
yes |
yes |
yes |
|
gain control: 50 dB |
no |
no |
yes |
yes |
yes |
yes |
|
angle steering for linear transducers |
no |
no |
±10° |
±10° |
±10° |
±30° |
|
real-time trace line with automatic calculation of spectrum parameters |
no |
no |
yes |
yes |
yes |
yes |
|
stereo sound: volume control |
no |
no |
yes |
yes |
yes |
yes |
|
PWD palette: 12 maps |
no |
no |
yes |
yes |
yes |
yes |
|
Doppler frequency selection: 2 frequencies / each transducer |
no | no |
yes |
yes |
yes |
yes |
|
File Formats image and video save/load |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
AVI, JPG, BMP, PNG, TIF |
yes |
yes |
yes |
yes |
yes |
yes |
|
DICOM formats: DCM (DICOM uncompressed), DCM (DICOM-JPEG RGB/YBR), DCM (DICOM-JPEG RGB/YBR Video) |
yes |
yes |
yes |
yes |
yes |
yes |
|
RAW data formats: TPD (Telemed Picture Data), TVD (Telemed Video Data) |
yes |
yes |
yes |
yes |
yes |
yes |
|
Ultrasound Information Technologies |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
report editor |
yes |
yes |
yes |
yes |
yes |
yes |
|
DICOM: Verification SCU, Modality Worklist (MWL) SCU, Modality Performed Procedure Step (MPPS) SCU, Store SCU (images, cines), Print SCU (grayscale, color) |
yes |
yes |
yes |
yes |
yes |
yes |
|
Image Management and Post-processing: RAW Data images/video review, processing, re-measure, save in any format. Import/export of images in any formats. |
yes |
yes |
yes |
yes |
yes |
yes |
|
direct e-mail sending with image or video attachment via Internet |
yes |
yes |
yes |
yes |
yes |
yes |
|
Remote control and Telemedicine |
yes |
yes |
yes |
yes |
yes |
yes |
|
Printing on system printer |
yes |
yes |
yes |
yes |
yes |
yes |
|
User Interface Customization |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
the set of predefined skin schemes for software interface |
yes |
yes |
yes |
yes |
yes |
yes |
|
the set of predefined buttons images |
yes |
yes |
yes |
yes |
yes |
yes |
|
Multilanguage support, languages: Chinese, English, Italian, Korean, Lithuanian, Magyar, Polish, Romanian, Russian, Spanish |
yes |
yes |
yes |
yes |
yes |
yes |
|
layout position, size |
B+M |
B+M |
B+M, B+PW |
B+M, B+PW |
B+M, B+PW |
B+M, B+PW |
|
ultrasound area size |
yes |
yes |
yes |
yes |
yes |
yes |
|
font size |
yes |
yes |
yes |
yes |
yes |
yes |
|
General Measurements and Calculations |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
B and Color Doppler mode general measurements and calculations: Distance, Length, Area and Circumference, Volume, Angle, Stenosis %, A/B Ratio |
yes |
yes |
yes |
yes |
yes |
yes |
|
M mode general measurements and calculations: Distance, Time, Velocity. Heart Rate, Stenosis %, A/B Ratio |
yes |
yes |
yes |
yes |
yes |
yes |
|
PW mode general measurements and calculations: Velocity, Pressure Gradient, Velocities difference, Pressure Gradients difference, Time interval, Acceleration, Resistivity Index, Heart Rate, Velocity minimum and maximum, Pressure Gradient minimum and maximum, Trace Time, Trace Velocity, Trace Pressure Gradient, Velocity Time Integral, Pulsatility Index, Velocities A/B Ratio, Pressure Gradients A/B Ratio, Velocity differences A/B Ratio, Pressure Gradient differences A/B Ratio, Time differences A/B Ratio, Accelerations A/B Ratio, Resistivity Indexes A/B Ratio, Velocity means A/B Ratio, Pressure Gradient means A/B Ratio, Pulsatility Indexes A/B Ratio, Velocity Time Integrals A/B Ratio |
no |
no |
yes |
yes |
yes |
yes |
|
PW mode Human Cardiology measurements and calculations |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
Left Ventricle, Mitral Valve, Aortic Valve, Right Ventricle, Tricuspid Valve, Pulmonic Valve, Pulmonary Vein, Shunts |
no |
no |
yes |
yes |
yes |
yes |
|
Human Measurements and Calculations Packages |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
General |
yes (excl. PW calc.) |
yes (excl. PW calc.) |
yes |
yes |
yes |
yes |
|
Obstetrics/Gynecology: Measurements, Estimated date of birth, Ratios, Estimated Fetal Weight and Average EFW, Gestational Age, Fetal Growth estimation, Growth Tables and Fetal Growth Curves, user-defined Growth Tables |
yes |
yes |
yes |
yes |
yes |
yes |
|
Gynecology, Abdominal, Urology, Endocrinology, Vascular, Cardiology |
yes |
yes |
yes |
yes |
yes |
yes |
|
Veterinary Calculations Packages |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
Canine OB, Feline OB, Ovine OB, Bovine OB, Equine OB, Llama OB, Goat OB, Animal Cardiology |
yes |
yes |
yes |
yes |
yes |
yes |
|
Ultrasound software |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
yes |
yes |
yes |
yes |
yes |
yes |
|
Recommended Computer Specification |
Echo Blaster 64 |
Echo Blaster 128 |
LogicScan 64 |
LogicScan 128 |
ClarUs |
SmartUs |
|
yes |
yes |
yes |
yes |
yes |
yes |
* depends on selected scan depth, scan angle, focus mode, High Line Density setting, computer speed